  • Abstracts - Problem of Transcutaneous Bilirubinometry
    2000 newborns were evaluated before and 2000 after the introduction of transcutaneous bilirubinometry with regard to frequency and methods of bilirubin determination The number of blood sampling procedures was reduced by 79 Transcutaneous bilirubinometry proves to be a safe method if correlation curves between the transcutaneous indices and the serum bilirubin of the individual nursery population are used together with set rules as to the indications for the controls by

    Original URL path: http://www.clinlab.ru/tm/bilirub/76.%20Schubiger.htm (2014-12-29)
    Open archived version from archive


  • Abstracts - Problem of Transcutaneous Bilirubinometry
    estimate sy X x 2 02 The device classified hyperbilirubinemia with a sensitivity of 94 and specificity of 77 5 A linear correlation existed between serum bilirubin values and the readings on the icterometer y 0 17x 1 74 r 63 sy X x 0 38 The device classified hyperbilirubinemia with a sensitivity of 82 and specificity of 74 The population studied was preselected for the presence of jaundice and

    Original URL path: http://www.clinlab.ru/tm/bilirub/77.%20Schumacher.htm (2014-12-29)
    Open archived version from archive

  • Abstracts - Problem of Transcutaneous Bilirubinometry
    and without the help of reference devices The establishment of an acceptable correlation between cutaneous and serum bilirubin requires 1 accurate measurement of serum bilirubin 2 accurate measurement of cutaneous bilirubin and 3 steady state conditions between the blood cutaneous bilirubin compartments The accuracy with which cutaneous bilirubin can be measured appears similar or better than that with which serum bilirubin is measured with older transcutaneous devices significant interobserver variability

    Original URL path: http://www.clinlab.ru/tm/bilirub/78.%20Schumacher.htm (2014-12-29)
    Open archived version from archive

  • Abstracts - Problem of Transcutaneous Bilirubinometry
    performed on cord blood Serial transcutaneous measurements in the first 72 hours of life in 63 ABO test positive group I and 31 ABO test negative group II infants were compared with a control group of 109 term ABO compatible infants group III During the first phase of this study the accuracy of TcBI was confirmed in 33 infants who had serum bilirubin determinations demonstrating a correlation coefficient of r

    Original URL path: http://www.clinlab.ru/tm/bilirub/79.%20Serrao.htm (2014-12-29)
    Open archived version from archive

  • Abstracts - Problem of Transcutaneous Bilirubinometry
    transcutaneous bilirubinometer was evaluated in 60 term and 10 preterm infants A significant correlation was found between the transcutaneous index and the total serum bilirubin concentration for both term and preterm infants The reliability of the transcutaneous bilirubinometer as a screening method was confirmed and index criteria for serum bilirubin analysis have been suggested for term babies The instrument was precise and accurate and the method both noninvasive and atraumatic

    Original URL path: http://www.clinlab.ru/tm/bilirub/81.%20Sheridan.htm (2014-12-29)
    Open archived version from archive

  • Abstracts - Problem of Transcutaneous Bilirubinometry
    the conclusion that the transcutaneous bilirubin index may be a valuable tool to help distinguish term babies with a total serum bilirubin value less than 13 mg dl from those with higher levels and that it serves as a valuable

    Original URL path: http://www.clinlab.ru/tm/bilirub/83.%20Strange.htm (2014-12-29)
    Open archived version from archive

  • Abstracts - Problem of Transcutaneous Bilirubinometry
    jaundice have blood samples sent to biochemistry laboratories for assessment of the plasma bilirubin concentration We planned to assess the cost in terms of finance medical staff time numbers of blood samples and family delay in leaving hospital We demonstrated that reflectance bilirubinometry is a reliable screening method for identifying which caucasian infants require to have plasma bilirubin concentrations measured in the laboratory The Minolta Airshields transcutaneous bilirubinometer provided reproducible

    Original URL path: http://www.clinlab.ru/tm/bilirub/84.%20Suckling.htm (2014-12-29)
    Open archived version from archive

  • Abstracts - Problem of Transcutaneous Bilirubinometry
    and compared with serum bilirubin levels measured by the American Optics A O bilirubinometer The serum bilirubin range was 3 8 18 4 mg dl The results showed that the TcB index correlates well with the serum bilirubin measurement the regression analysis was r 0 878 y 10 78 0 909 x P less than 0 001 However the regression line showed that at a serum bilirubin level above 12

    Original URL path: http://www.clinlab.ru/tm/bilirub/85.%20Taha.htm (2014-12-29)